2. Installation Manual
- 35 -
(4) Checking the Sensor Serial No.
1) Purpose
If the sensor serial number and the sensor serial number stored in the flash ROM of the
PWB-50Di differ from the image data file name stored in the hard drive of the Control PC
due to replacing the PWB-50Di or imaging unit, the connected sensor can not be detected
after the CXDI application is launched. In that case an error message appears.
2) Notes
2-1) Check the sensor serial No. whenever:
a. PWB-50Di is replaced.
b. Sensor in the imaging unit is replaced.
2-2) This checking procedure must be performed with the Control PC, Imaging Unit and
all the other equipments connected and started up.
3) Procedure
3-1)
Check the sensor serial No. by referring to “Sensor serial number setting” in the
Tool Software Operation Manual for Ethernet, and write the number as required.