2. Installation Manual
- 26 -
5.5 Setting
the
Fixed ROI Areas
1) Purpose
Set the fixed ROI area on the sensor to expose by the fixed ROI area because user can
not get the proper image by the Auto ROI area.
2) Setting
method
2-1) Investigate the actual size and position of the ROI that is required.
2-2) Designate the SIZE, POSITION, and NUMBER (max 5)* on the sensor.
* However, in the case of using the new function “Display of AEC (Automated Exposure Control)
Field in Preview Screen” added from CXDI System Software Ver.6.2, NUMBER that can be
specified is max 3.
Ws=350mm
H
s=
430mm
(0,0)
W
H
Sensor
X
Y
2-3) Convert the size and position of the ROIs in 2) to pixel values. The pixel size of the
sensor is 160µm. For multiple values, use X’, Y’, W’, H’, X”, Y”, W”, and H” for
calculations.
X/160 µm
Let this value equal A
Y/160 µm
Let this value equal B
W/160 µm
Let this value equal C
H/160 µm
Let this value equal D
2-4) Open the “DENSITY ADJUSTMENT CONTROL” screen when the normal
imaging screen is active.
SYSTEM
Æ
SETUP MENU
Æ
SYSTEM SETTINGS
Æ
DENSITY
ADJUSTMENT CONTROL
Summary of Contents for CXDI-50C
Page 15: ......
Page 23: ...2 Installation Manual 6 3 Power Box assemble package 4 Grid optional ...
Page 35: ...2 Installation Manual 18 When normal imaging When timeout due to RX_REQ not negating ...
Page 63: ...2 Installation Manual 46 Fig 3 Input parameters ...
Page 115: ...2 Installation Manual 98 2 Power Box Unit mm Mass 4 2Kg ...
Page 118: ......
Page 163: ......
Page 167: ...CXDI SYSTEM 1 2 Rev 04 Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Canon Inc ...
Page 186: ......
Page 188: ......
Page 189: ...Content 1 Gathering Information about Problems 1 2 How to back up and recover the system 3 ...