2. Installation Manual
- 76 -
<
Reference 2>
Example when the graph does not form a straight line
In the data and graph below, the data was obtained by using the Kodak MLP 190 to print
out the SMPTE test image, and the image data was measured.
(1): Shows normal data.
(2): Shows the case when the CXDI printer parameters are not set linearly.
(3): Shows the case when the printer was not calibrated properly.
SMPTE Density
Step
1
2
3
4
5
6
7
8
9
10
11
Percentage
0
10
20
30
40
50
60
70
80
90
100
0.17 0.46 0.75 1.03 1.36 1.66 1.98 2.29
2.59
2.90
3.20
(2)
Inadequate parameter settings
0.17 0.29 0.45 0.61 0.80 1.02 1.26 1.59
2.02
2.53
3.19
(3) Inadequate
0.17 0.33 0.55 0.81 1.08 1.36 1.66 1.97
2.28
2.59
2.94
(1) Normal
0.00
0.50
1.00
1.50
2.00
2.50
3.00
3.50
1
2
3
4
5
6
7
8
9
10
11
Step
Densi
ty
Normal
Inadequate
setting
Inadequate
* For improper parameters, the printer parameter “-S CS000” was not entered.
* For the inadequate calibration, the printer calibration data was set too low.
Summary of Contents for CXDI-50C
Page 15: ......
Page 23: ...2 Installation Manual 6 3 Power Box assemble package 4 Grid optional ...
Page 35: ...2 Installation Manual 18 When normal imaging When timeout due to RX_REQ not negating ...
Page 63: ...2 Installation Manual 46 Fig 3 Input parameters ...
Page 115: ...2 Installation Manual 98 2 Power Box Unit mm Mass 4 2Kg ...
Page 118: ......
Page 163: ......
Page 167: ...CXDI SYSTEM 1 2 Rev 04 Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Canon Inc ...
Page 186: ......
Page 188: ......
Page 189: ...Content 1 Gathering Information about Problems 1 2 How to back up and recover the system 3 ...