3-8) Check the communication test in the sequence below to verify whether the CXDI is
now part of the network. To check the connections at the TCP/IP level, use the “ping”
command from the command prompt.
Start
Æ
Programs
Æ
Command Prompt
When the IP address of the connection destination is “173.17.7.123,” for instance, the
following messages will be repeated.
z
If the CXDI has been connected properly:
Pic:>ping 172.17.7.123 (input on the DOS screen)
Pinging 17217.7.123 With 32 bytes of data:
Reply from 172.17.7.123:bytes=32 time <10ms TTL=255
Reply from 172.17.7.123:bytes=32 time <10ms TTL=255
Reply from 172.17.7.123:bytes=32 time <10ms TTL=255
Reply from 172.17.7.123:bytes=32 time <10ms TTL=2550
z
If the CXDI has not been connected properly:
Pic:> ping 172.17.7.123 (input on the DOS screen)
Pinging 17217.7.123 With 32 bytes of data:
Request time out
Request time out
Request time out
Request time out