2. Installation
- 12 -
4.2 Affixing the Insulation Sheets
This section describes the affixing positions and affixing procedure of the insulation sheets
(accessories) used to electrically isolate the Bucky unit and imaging unit when an imaging unit is
mounted on the Bucky unit.
The affixing positions of the insulation sheets in this manual are based on installation of the
imaging unit on the Liebel-Flarsheim Bucky unit rail.
*Notes
1. If installing an imaging unit to a Bucky unit other than a Liebel-Flarsheim Bucky unit,
stand or table, obtain installation sheets from the distributor, and affix the insulation
sheets to ensure that the Bucky unit, stand or table is electrically isolated from the
imaging unit.
2. The parts used to secure the imaging unit to the Bucky unit, stand or table must be
obtained from the distributor. The securing parts (including the mounting screws)
must also be electrically insulated by insulation sheets or other means to ensure that
the imaging unit is electrically isolated.
3. After installation of the imaging unit in the Bucky unit, stand or cable is complete, use
a tester or other instrument to check that the imaging unit is electrically isolated.
4.2.2 Insulation Sheets Affixing Procedure (Using the attached insulation sheets)
(1) Remove any oils from the surface where the insulation sheets will be affixed on the imaging
unit (grounding surface with the Bucky unit).
(2) Affix the four (2) insulation sheets to the (1) imaging unit rear side.
Note: Avoid making bent or raised sections when affixing.
*Enlarged View of DT3
(Af
fi
x b
y
alig
nin
g with
th
e
radi
us e
dge
o
f
th
e f
rame
)
Frame
(Affix by aligning with the
radius edge of the frame)
* Affix to the
rear surface
of the frame
DT3
Summary of Contents for CXDI-40G COMPACT
Page 24: ...2 Installation 4 2 CXDI SYSTEM III Assembly Packaging Accessories Box Power Box ...
Page 56: ...2 Installation 36 8 3 When Timeout due to RX_REQ not Negating ...
Page 135: ...2 Installation 115 16 2 Power Box Mass 4 2 Kg not including cable Unit mm ...
Page 136: ...2 Installation 116 16 3 Ready Lamp Unit Mass 0 5 Kg Unit mm ...