12.14 Backing Up When Installing
1) Purpose
In case of re-installing the CXDI application, the necessary files ex) the exposure
position and other parameters must be backed up so that can be restored at the status of
first installation.
2) Necessary
items
2-1) Removable drive such as MO drive, USB memory or external HDD that can connect
to USB 2.0
3) Notes
3-1) Before performing backup procedure, delete any “BodyPart” and image data exposed
for tests.
Deleting image data: Refer to the “Deleting Data” item.
Deleting “BodyPart”: Refer to the CXDI Series Operation Manual.
3-2) Backup should be performed immediately before handing over the product to the
customer only for new installations.
4) Connections
4-1) Check that all equipment is turned off.
4-2) Connect the keyboard and mouse removal drive to the control PC.
5) Settings
5-1) Turn on the operation unit power and then the control PC power.
5-2) The CXDI application starts. Press the [Alt] + [Tab] keys to switch the program to the
Command Prompt screen.
5-3) The message “Welcome to Canon CXDI” appears. Input [8] and press the [Enter] key.
(Select “8 Exit”.)
5-4) The Windows XP desktop screen appears.
5-5) Right click on My Computer on the Desktop screen, and select Explorer from Menu.
5-6) Explorer appears. Copy the CCR folder onto Removable Drive.
Summary of Contents for CXDI-40G COMPACT
Page 24: ...2 Installation 4 2 CXDI SYSTEM III Assembly Packaging Accessories Box Power Box ...
Page 56: ...2 Installation 36 8 3 When Timeout due to RX_REQ not Negating ...
Page 135: ...2 Installation 115 16 2 Power Box Mass 4 2 Kg not including cable Unit mm ...
Page 136: ...2 Installation 116 16 3 Ready Lamp Unit Mass 0 5 Kg Unit mm ...