2. Installation
- 26 -
●
Using a measure or rope, align the vertical position of the sensor unit to the center of the tube so
that the distance from the tube to the top and bottom side of the sensor is symmetric.
[Side View]
(3) Turn on the irradiation ramp of the X-ray generator, cross line of the X-ray tube is cast over the
sensor unit. Align the position to meet the requirements described in 1) and 2) below.
1) Align the cross line of X-ray tube with the cross line of the sensor unit. If the position is mismatched
as shown in [NG] below, the sensor unit and the stand are not horizontal. In this case, put the spacer
between the stand base and the floor to align the slight differences.
Sensor Unit
Stand
Floor
Spacer
x
[OK]
[NG]