![Canon CXDI-40G COMPACT Service Manual Download Page 158](http://html1.mh-extra.com/html/canon/cxdi-40g-compact/cxdi-40g-compact_service-manual_4002850158.webp)
4. Repair Guide
- 4 -
Sensor Interface Cable Wiring
Adjust the length of the sensor interface cable when you replace it.
Error!
(2) Assembly
Assembling is reverse of disassembling basically.
Flat cable connector is not quite resistant to wear. (The reliability of the communication will
decrease). As the cable manufacturer recommends, flat cables should be exchanged once
removed.
However, the flat cable between PCB-40EAD and PCB-D can be continuously used after it is
removed for the PCB-40EAD replacement considering the service’s profitability. If any
problems arise after the replacement, check the possible failure parts and take appropriate
action.
Note: Connect and disconnect the connector of the flat cable carefully. Connection and
disconnection of the cable is allowed up to three times.
Adjust the cable length here
Continuous use is
allowed for this part