14. Image Quality
1) Purpose
To check the CXDI final image quality
To check that the images are not affected by extraneous magnetic noise (emanating from
CT or MRI apparatus, CRT monitors and other equipment that generate magnetic fields
near the sensor unit)
2) Purpose
(1) Phantom
(2) High-resolution monitor or DICOM printer
<Procedure>
Procedure/Item Operation/Conditions
1. Adjust the distance between the focal point of the X-
ray tube and the CXDI sensor unit.
X cm: Differs according to the grid being used.
2. Perform calibration
SYSTEM > CALIBRATION
Perform calibration and note that phototimer and
other parts except for grid that are located in the X-
ray entrance surface must be installed in advance.
Do not install the grid.
3. Install the grid.
Check the grid installation angle.
X-ray tube focal
point
X cm
Right angle at
the center
CXDI Sensor Unit
Install the grid
[Positioning]
Position the X-ray tube and
CXDI Sensor Unit
Calibration
Summary of Contents for CXDI-40G COMPACT
Page 24: ...2 Installation 4 2 CXDI SYSTEM III Assembly Packaging Accessories Box Power Box ...
Page 56: ...2 Installation 36 8 3 When Timeout due to RX_REQ not Negating ...
Page 135: ...2 Installation 115 16 2 Power Box Mass 4 2 Kg not including cable Unit mm ...
Page 136: ...2 Installation 116 16 3 Ready Lamp Unit Mass 0 5 Kg Unit mm ...