2. Installation
- 48 -
11. Setting the Fixed ROI Areas
(1) Purpose
Set the fixed ROI area on the sensor to expose by the fixed ROI area because user cannot get the
proper image by the Auto ROI area.
(2) Setting
method
1) Investigate the actual size and position of the ROI that is required.
2) Designate the SIZE, POSITION, and NUMBER (max 5) on the sensor.
* However, in the case of using the new function “Display of AEC (Automated Exposure
Control) Field in Preview Screen” added from CXDI System Software Ver.6.2, NUMBER
that can be specified is max 3.
Ws=430mm
H
s=430
mm
(0,0)
W
H
Sensor
X
Y
3) Convert the size and position of the ROIs in 2) to pixel values. The pixel size of the
sensor is 160µm. For multiple values, use X’, Y’, W’, H’, X”, Y”, W”, and H” for
calculations.
X/160 µm
Let this value equal A
Y/160 µm
Let this value equal B
W/160 µm
Let this value equal C
H/160 µm
Let this value equal D
4) Open the “DENSITY ADJUSTMENT CONTROL” screen when the normal imaging
screen is active.
SYSTEM
Æ
SETUP MENU
Æ
SYSTEM SETTINGS
Æ
DENSITY
ADJUSTMENT CONTROL
Summary of Contents for CXDI-40G COMPACT
Page 24: ...2 Installation 4 2 CXDI SYSTEM III Assembly Packaging Accessories Box Power Box ...
Page 56: ...2 Installation 36 8 3 When Timeout due to RX_REQ not Negating ...
Page 135: ...2 Installation 115 16 2 Power Box Mass 4 2 Kg not including cable Unit mm ...
Page 136: ...2 Installation 116 16 3 Ready Lamp Unit Mass 0 5 Kg Unit mm ...