![Canon CXDI-40G COMPACT Service Manual Download Page 15](http://html1.mh-extra.com/html/canon/cxdi-40g-compact/cxdi-40g-compact_service-manual_4002850015.webp)
1. General
- 8 -
5. Specifications
The CXDI-40G COMPACT has a secondary scanning drive unit, data reading unit, and A/D
conversion unit on one side of the LANMIT, and it is a sensor unit that reads image data from one
direction for a single panel.
The X-ray image is converted to photoelectric signals by the sensor unit, and the scanned electrical
signals (image signals) undergo A/D conversion, pass through the power box, and are transferred to
the control PC over an Ethernet cable.
Item
40G COMPACT
Remarks
Operational format
Upright stand, table, universal stand
Installation on a motor vehicle
is not possible.
Effective exposure range
430 x 430 mm
Total number of pixels
2706 x 2700 pixels
Number of effective pixels
7.2 million (2688 x 2688 pixels)
Pixel pitch
160
µ
m x 160
µ
m
Fluorescent substance
GOS
LANMIT sensor
Sensor sensitivity
Same as CXDI-40EG
Exposure time
0 ms to 1 sec., 1 to 3 sec.
Output gradations
12 bits (4096 gradations), A/D 14 bits
Transfer method
Ethernet between sensor unit and PC (via
power box)
Preview time
Approx. 3 sec
Exposure cycle time
15 sec. (typ.)
External dimensions (Sensor
unit)
470(W)×548(L)×32(H) mm
Not including insulation
sheet
External dimensions (Power
box)
358(W)×212(L)×75(H) mm
Color of sensor unit exterior
Black (Cathodic electrodeposition coating)
Mass (Sensor unit)
11.0 kg
Not including cables or
ready lamp
Mass (Power box)
4.2 kg
Distance from exterior CFRP
surface to sensor surface (glass
surface)
6.6±0.5 mm
Patient contact surface
None
Exterior strength (imaging
unit)
Excess local weight: 12 kg (118 N) with
15 mm diameter for 1 minute
Incoming X-ray surface
only
Exterior strength (imaging
unit, power box)
Excess local weight: 20 kg (196 N) with
15 mm diameter for 1 minute
Locations other than
incoming X-ray surface
Summary of Contents for CXDI-40G COMPACT
Page 24: ...2 Installation 4 2 CXDI SYSTEM III Assembly Packaging Accessories Box Power Box ...
Page 56: ...2 Installation 36 8 3 When Timeout due to RX_REQ not Negating ...
Page 135: ...2 Installation 115 16 2 Power Box Mass 4 2 Kg not including cable Unit mm ...
Page 136: ...2 Installation 116 16 3 Ready Lamp Unit Mass 0 5 Kg Unit mm ...