![Canon CXDI-40G COMPACT Service Manual Download Page 64](http://html1.mh-extra.com/html/canon/cxdi-40g-compact/cxdi-40g-compact_service-manual_4002850064.webp)
2. Installation
- 44 -
9.3.4 Connection Setup for a Multiple Number of Sensor Units
When the control PC is to be connected to a multiple number (2 or more) of sensor units, it is
recommended that the equipment be connected via a hub using straight cables (supporting
category 5 or higher). You can use both straight and crossover cables when using a hub that
provides Auto MDI/Auto MDI-X function.
Use of units other than a hub, such as a dual-port LAN card or multiple number of LAN cards,
is not recommended (since no guarantees are made for operation).
When the control PC is to be connected to a single sensor unit, connect the equipment using a
crossover cable as in the past.
* Note:
Bear in mind that when connecting the equipment to a hospital LAN, for instance, if the
environment is not a favorable one for Ethernet (there is too much traffic or high noise levels),
the equipment may not be able to demonstrate its performance to the full. Therefore, the
construction of a network consisting solely of the CXDI system is strongly recommended.
*Up to three identical sensors can be connected. Each sensor must be paired with a power box.
*The LAN cables and hub are procured from the sales companies (or distributor).
Connect as shown below if a multiple number of sensor units are to be connected. Example:
CXDI-40G COMPACT
+ CXDI-50G + control PC
9.3.4.1 General Flow for Setup Procedure
1)Proceed with the system connections between the single sensor unit and control PC.
2) Change the IP address settings.
3) Check the IP address and subnet mask settings.
4) Perform steps 1) to 3) for every sensor to be connected.
5) Edit the screwcap.ini file.
6) Start up the system software.
7) Set the CCR console menu.
8) Reboot the system software.
CXDI-50G
Power Box
Power Box
HUB
Control
PC
CXDI-40G COMPACT
Summary of Contents for CXDI-40G COMPACT
Page 24: ...2 Installation 4 2 CXDI SYSTEM III Assembly Packaging Accessories Box Power Box ...
Page 56: ...2 Installation 36 8 3 When Timeout due to RX_REQ not Negating ...
Page 135: ...2 Installation 115 16 2 Power Box Mass 4 2 Kg not including cable Unit mm ...
Page 136: ...2 Installation 116 16 3 Ready Lamp Unit Mass 0 5 Kg Unit mm ...