![Canon CXDI-40G COMPACT Service Manual Download Page 48](http://html1.mh-extra.com/html/canon/cxdi-40g-compact/cxdi-40g-compact_service-manual_4002850048.webp)
2. Installation
- 28 -
5.3.2 Table
(1) Fix the mirror (first-face mirror) at the center of the sensor unit.
(2) Place the level on the sensor unit. Align the table and the sensor horizontally by the adjuster of the
table or the spacer.
(3) Adjust the height of the table to space the appropriate exposure distance.
Distance from the tube to the sensor unit should be 150cm.
●
Using a measure or rope, align the vertical and horizontal position of the sensor unit to the center of
the tube so that the distance from the tube to side to side and up and down of the sensor is
symmetric.
Level
Adjuster
x
Fix the mirror