12.7 Entering Control PC Serial Number
1) Purpose
Set the product serial number (Control PC) to the “Device Serial Number” of the
“DICOM header”.
2) Procedure
2-1) Start up the CXDI system.
2-2) After the exposure screen appears on the operation unit, use the keyboard to enter
Debug mode. (Use [Alt] + [Tab].)
2-3) “Welcome to CCR” screen appears. Select the command “1. Set-Up…”
2-4) The “Setting Mode (0: Normal, 1: Expert) [0=0
×
0]:” is prompted. Select “0: Normal”
2-5) “CCR SETUP MENU” appears. Select the command ”1. System Setup”.
2-6) “CCR Serial Number [0=0
×
0] : ” appears. Enter the six-digit number indicated on the
naming label of the control PC unit. Press [Enter] key until “CCR SETUP MENU”
appears. [Fig 1]
[
[
F
F
i
i
g
g
1
1
]
]
2-7) Press [Esc] key after “CCR SETUP MENU” appears to return to “Welcome to CCR”
screen.
2-8) Select “8-Exit”to exit CXDI application software.
2-9) This returns you to the Windows NT desktop. Restart the CXDI application, and
perform the procedure from steps 2) to 5). Check that the serial number for the “CCR
Serial Number” item was entered correctly in step 6).
* Restart the CXDI application. The screen displays the following message:
Alert System Info Error (-6) A/D board info is updated. Click “OK”
Enter here
Summary of Contents for CXDI-40G COMPACT
Page 24: ...2 Installation 4 2 CXDI SYSTEM III Assembly Packaging Accessories Box Power Box ...
Page 56: ...2 Installation 36 8 3 When Timeout due to RX_REQ not Negating ...
Page 135: ...2 Installation 115 16 2 Power Box Mass 4 2 Kg not including cable Unit mm ...
Page 136: ...2 Installation 116 16 3 Ready Lamp Unit Mass 0 5 Kg Unit mm ...