According to “IEC60601-1-1:2000”, devices installed in the patient environment are restricted
to “electric medical devices conforming to IEC60601-1”.
The Control PC and operation unit are classified under the data processing device standard
(IEC60950), therefore these items should not be installed in the patient environment.
The patient environment described below is an example cited from “IEC60601-1-1:2000” – the
measurements are only guidelines. However, the “IEC60601-1-1:2000” example must be treated
as the standard.
Therefore, the Control PC and operation unit must be installed in a location further than the
measurements below (outside of the patient environment).
*Areas where the patient moves (not only during imaging but when entering and leaving
the room, etc.) are also considered as part of the patient environment, therefore the
installation location should be determined upon consultation with the user regarding
areas outside of the patient environment.
Example of patient environment
Note: These measurements are only guidelines.
Summary of Contents for CXDI-40G COMPACT
Page 24: ...2 Installation 4 2 CXDI SYSTEM III Assembly Packaging Accessories Box Power Box ...
Page 56: ...2 Installation 36 8 3 When Timeout due to RX_REQ not Negating ...
Page 135: ...2 Installation 115 16 2 Power Box Mass 4 2 Kg not including cable Unit mm ...
Page 136: ...2 Installation 116 16 3 Ready Lamp Unit Mass 0 5 Kg Unit mm ...