CXDI-40G COMPACT Inspection Check Sheet
< I n s p e c t i o n I t e m s >
ITEM Category NO Item Characteristics values
Test means
Record
mA
settings
which calibration is
implemented must be
noted
msec
Image quality
2
Self-diagnosis
pass/fail
Tests 1 to 5 must be
passed.
Visual check
TEST1
TEST2
TEST3
TEST4
TEST5
8 Image
Quality
1
Phantom
exposure
There must be no
artifacts, shading, grid
stripes, etc.
Visual check
9
Transmission test
1
Dcap, 200 times Transmission must be
successful.
Dcap.exe
10 Ping
test 1
Ping, 100,000
times
Communication must be
successful.
Pingtest.exe
1
Dirt, dust inside
PC
There must be no dirt or
dust
Visual check
2
Event logs
Application logs
Watson logs
There must be no errors
(excluding errors which
have been dealt with).
Visual check
3
System logs
OPU3 logs
There must be no errors
(excluding errors which
have been dealt with).
Visual check
4 Touch
panel
There must be no
deviation or other
problems.
Visual check
5 Date,
time
The date and time must
be correct.
Visual check
6
CCR application
version
- Visual
check
Ver.
11 PC-related
items
7
CCR folder
backup
-
External media,
etc.
12 System
exterior 1
Dirt
There must be no dirt.
Visual check
1
Hookup with RIS,
patient data
terminals
It must be possible to
transmit and receive the
data correctly.
Visual check
2
Exposure
information,
patient data in
film server
There must be no errors
in the information or
data.
Visual check
13 Final
checks
3
X-ray generator,
RIS and other
equipment
The equipment must be
the same as before the
inspection work.
Visual check
Summary of Contents for CXDI-40G COMPACT
Page 24: ...2 Installation 4 2 CXDI SYSTEM III Assembly Packaging Accessories Box Power Box ...
Page 56: ...2 Installation 36 8 3 When Timeout due to RX_REQ not Negating ...
Page 135: ...2 Installation 115 16 2 Power Box Mass 4 2 Kg not including cable Unit mm ...
Page 136: ...2 Installation 116 16 3 Ready Lamp Unit Mass 0 5 Kg Unit mm ...