2. Installation
- 16 -
3) After passing the ready lamp cable through the edge saddle, use two cable ties to secure the
cable to the cable clamps (two locations).
4) Loosen the nut of the sensor cable, lay the sensor cable on the cable mounting plate, tighten the
sensor cable nut; then securely tighten the conductive holder over the copper foil covering.
Note
Loosen a SKINTOP spiral of the sensor cable so that the sensor cable connector can be
connected to the connector of the imaging unit as needed.
Route the sensor cable while taking care that excessive force is not applied to it.
To prevent damage in case the cable is pulled by a strong force, adjust by tightening a
SKINTOP spiral as shown in the figure below.
Cable clamps
Edge saddle
Nut for the sensor cable
Sensor cable
Conductive holder