CXDI-40G COMPACT Inspection Check Sheet
< I n s p e c t i o n I t e m s >
ITEM Category NO Item Characteristics values
Test means
Record
1 Sensor
cable
The cable must not be
flattened or twisted and
its covering must not be
damaged.
Visual check
2
Sensor cable
connector
The connector must be
connected securely,
and there must be no
play.
Visual check,
touch
3 Ready
Lamp
The ready lamp must
not be flattened or
twisted and its covering
must not be damaged.
Visual check
4
Ready Lamp
Connector
The connector must be
connected securely.
Visual check,
touch
5 Firmware
version
-
Visual
check
Ver
6
PLD code version
-
Visual check
Ver
1 Imaging
unit
7
Imaging unit IP
address
- Visual
check
1 Connector
The connector must be
connected securely.
Visual check,
touch
2 Power
box
2 Power
switch
It must be possible to
set this switch to its ON
and OFF positions
properly.
Visual check,
touch
3
Control PC
1 Connector
The connector must be
connected securely.
Visual check
1 Startup
The system power must
come on when the
power of the control PC
is turned on..
Visual check
2 CCR
startup
The system must start
up with no errors.
3 Status
lamps
The lamps must provide
the correct indication
that corresponds to the
sensor statuses.
4
System connections
4 Shutdown
The system power must
go off when the power
of the control PC is
turned off.
1
X-ray generator
settings
The conditions under
which calibration is
implemented must be
noted..
Visual check
2
Preparation for
exposure
Phototimer must be
installed. Grid must not
be installed.
Visual check
5
Calibration
3
Calibration
processing
“Calibration completed
successfully.”
Visual check
6 Grid
installation
1
Grid installation
angle check
Within ±1.0 degree
Audit_grid.exe
7
Self-diagnosis
1
X-ray generator
The conditions under
Visual check
kV
Summary of Contents for CXDI-40G COMPACT
Page 24: ...2 Installation 4 2 CXDI SYSTEM III Assembly Packaging Accessories Box Power Box ...
Page 56: ...2 Installation 36 8 3 When Timeout due to RX_REQ not Negating ...
Page 135: ...2 Installation 115 16 2 Power Box Mass 4 2 Kg not including cable Unit mm ...
Page 136: ...2 Installation 116 16 3 Ready Lamp Unit Mass 0 5 Kg Unit mm ...