Flow Installation Parameters
31
30
29
28
27
26
25
24
23
22
21
20
19
18
17
16
—
MT4
MT3
MT2
MT1
MT0 TOR
TLO
TT1
TT0
EXT
—
AUS
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
—
WPH
dM3 dM2
dM1
dM0
—
DM3 DM2 DM1 DM0
Value
dM3-dM0: Plate (d) Metal Type
DM3-DM0: Pipe (D) Metal Type
0
Zero Thermal Expansion
1
Generic Carbon Steel
2
Generic 300-Series Stainless Steel
3
304, 304H Stainless Steel (ASTM A312-304)
4
316, 316H Stainless Steel (ASTM A312-316)
5
Monel and Related Nickel Alloys
6
Monel 400
7
Yellow Brass (ASTM B36, B134, B135)
8
Inconel-X, Annealed
9
Pure Nickel
10
Hastelloy C-22
11
Titanium, 20 °C to 100 °C
Value
AUS: Expansion Coefficient Source
0
Coefficients are based on SI Tables
1
Coefficients are based on US Customary Tables
Value
WPH: Weep Hole Installed
0
No weep hole
1
Weep hole installed
Value
TT1-TT0: Tap Type
0
Corner
1
Flange
2
D and D2
Value
TLO: Static Tap Location
0
Upstream
1
Downstream
Value
TOR: Tap Orientation
0
deg90 (Eccentric Orifice Only)
1
deg180
Value
EXT: Extended Temperature Range
0
Fixed Alpha
1
Alpha corrected to higher temperatures
Value
MT4-MT0: Meter Type Information
Meter Type: Classical Herschell Venturi
0
Vcalibrated
1 – 3
—
33
Scanner 3100 Modbus Protocol
Section 1