Code
Description
Arg 1
Arg 2
Arg 3
Arg 4
Config
Lock
If Config Lock = Yes, an active configuration lock in the Scanner prevents changes to the register.
610000 Sends the selected
Data Set to the
selected Slave
Device. This will
cause the registers
within the Scanner
3100 to be written to
the configuration of
the connected slave.
The configuration
change may take
up to 15 seconds to
complete.
0 = None
1 - 20 = Slave1 - Slave20
To synchronize a
slave configuration
change made in the
Scanner 3100 with
the corresponding
slave device, write
the desired value
from the list below
to Argument 2, and
write command. To
confirm the change is
accepted by the slave
device, read register
5056.
0 = None
1 = Device Name
2 = Archive
Configuration
3 = Flow Run
Configuration
4 = Flow Run
Maintenance
5 = Gas Composition
6 = Cone Calibration
7 = Turbine Input 1
Configuration
8 = Turbine Input 1
K-Factor Calibration
9 = Turbine Input 2
Configuration
10 = Turbine Input 2
K-Factor Calibration
11 = Diff Pressure
Configuration
12 = Diff Pressure
Calibration
13 = Static Pressure
Configuration
14 = Static Pressure
Calibration
15 = Temperature
Configuration
16 = Temperature
Calibration
17 = Analog Input 1
Configuration / PID
Controller Settings
18 = Analog Input 1
Calibration
19 = Analog Input 2
Configuration
20 = Analog Input 2
Calibration
—
—
Yes
15
Scanner 3100 Modbus Protocol
Section 1