10.2.2 Measurement Calibration
All measurement calibrations are performed by adjusting the measurement reading up or
down using a calibration coefficient. The coefficient value ranges from -127 (adjust full-scale
downward, zero upward) to + 127 (adjust full-scale upward, zero downward). When used in
combination with the MEAS commands, an automated measurement calibration procedure
can be implemented by adjusting the coefficient and reading the new measurement value
iteratively until the delta between an external reference measurement device and the AC
source internal measurement is within the specified accuracy.
CALibrate:MEASure:CURRent[:FSCale] <NRf>
This command affects the calibration of the AC current measurement at full scale.
Command Syntax
CALibrate:MEASure:CURRent[:FSCale] <NRf>
Parameters
<NRf> (value range -127 to +127)
Examples CAL:MEAS:CURR
-23
Query Syntax
CALibrate:MEASure:CURRent?
Returned Parameters
<NR1> (value range -127 to +127)
Related Commands
CALibrate:MEASure:VOLTage
CALibrate:MEASure:CURRent:ZERO <NRf>
This command affects the calibration of the AC current measurement zero offset.
Command Syntax
CALibrate:MEASure:CURRent:ZERO <NRf>
Parameters
<NRf> (value range -127 to +127)
Examples CAL:MEAS:CURR:ZERO
+12
Query Syntax
CALibrate:MEASure:CURRent:ZERO?
Returned Parameters
<NR1> (value range -127 to +127)
Related Commands
CALibrate:MEASure:CURRent
CALibrate:MEASure:PCURrent[:FSCale] <NRf>
This command affects the calibration of the AC peak current measurement at full scale.
Command Syntax
CALibrate:MEASure:CURRent[:FSCale] <NRf>
Parameters
<NRf> (value range -127 to +127)
Examples CAL:MEAS:CURR
-23
Query Syntax
CALibrate:MEASure:CURRent?
Returned Parameters
<NR1> (value range -127 to +127)
Related Commands
CALibrate:MEASure:VOLTage
CALibrate:MEASure:PCURrent:ZERO <NRf>
This command affects the calibration of the AC current measurement zero offset.
Command Syntax
CALibrate:MEASure:CURRent:ZERO <NRf>
Parameters
<NRf> (value range -127 to +127)
Examples CAL:MEAS:CURR:ZERO
+12
Query Syntax
CALibrate:MEASure:CURRent:ZERO?
Returned Parameters
<NR1> (value range -127 to +127)
Related Commands
CALibrate:MEASure:CURRent
84 April
2007
2001RP
Summary of Contents for 2001RP
Page 33: ...User and Programming Manual Rev N Figure 3 7 Test Setup 2001RP 24 April 2007 2001RP...
Page 47: ......
Page 60: ...User and Programming Manual Rev N Figure 6 3 Internal Adjustments 2001RP April 2007 51...
Page 65: ...User and Programming Manual Rev N Figure 7 1 Assembly Location 56 April 2007 2001RP...
Page 68: ...User and Programming Manual Rev N Figure 7 2 Replaceable Fuse Locations 2001RP April 2007 59...
Page 70: ...User and Programming Manual Rev N Figure 8 1 Rear Panel View 2001RP April 2007 61...
Page 90: ......