User and Programming Manual - Rev N
6.4.3 Open Sense Adjustment
Connect the test equipment as shown in Figure 6-1.
NOTE:
The 2001RP must be configured for the Constant Current Mode of
operation for this procedure. For this mode jumpers W1 and W2 on
the controller must be removed.
Disconnect the SENSE HIGH terminal on the rear panel terminal strip. Connect the external
AC DMM between the two output POWER terminals on the rear panel.
Program the output to 100.0 volts and adjust R2 on the Mother Board for an output voltage
of 115 volts.
Turn off the front panel circuit breaker and reinstall the top cover. Also reconnect the
external SENSE terminals at the rear panel.