User and Programming Manual - Rev N
10.9.1 *CLS
This command clears the following registers (see chapter 12 for descriptions of all status
registers):
•
Standard Event Status
•
Status Byte
•
Error Queue
Command Syntax
*CLS
Parameters None
10.9.2 *ESE
This command programs the Standard Event Status Enable register bits. The programming
determines which events of the Standard Event Status Event register (see *ESR?) are
allowed to set the ESB (Event Summary Bit) of the Status Byte register. A "1" in the bit
position enables the corresponding event. All of the enabled events of the Standard Event
Status Enable Register are logically OR-ed to cause the Event Summary Bit (ESB) of the
Status Byte Register to be set. See Section 16 for descriptions of the Standard Event Status
registers.
The query reads the Standard Event Status Enable register.
Table 10-2: Bit Configuration of Standard Event Status Enable Register
Bit Position
7
6
5
4
3
2
1
0
Bit Name
PON
not
used
CME EXE DDE QYE not
used
not
used
Bit
Weight 128 32
16 8 4
CME
Command error
DDE
Device-dependent error
EXE
Execution error
QYE
Query error
PON Power-on
Command Syntax
*ESE <NRf>
Parameters
0 - 255
Power-On Value
128
Example *ESE
129
Query Syntax
*ESE?
Returned Parameters
<NR1>(Register value)
Related Commands
*ESR? *STB?
2001RP April
2007
103
Summary of Contents for 2001RP
Page 33: ...User and Programming Manual Rev N Figure 3 7 Test Setup 2001RP 24 April 2007 2001RP...
Page 47: ......
Page 60: ...User and Programming Manual Rev N Figure 6 3 Internal Adjustments 2001RP April 2007 51...
Page 65: ...User and Programming Manual Rev N Figure 7 1 Assembly Location 56 April 2007 2001RP...
Page 68: ...User and Programming Manual Rev N Figure 7 2 Replaceable Fuse Locations 2001RP April 2007 59...
Page 70: ...User and Programming Manual Rev N Figure 8 1 Rear Panel View 2001RP April 2007 61...