[SOURce:]VOLT:RANGe:AUTO
This command sets the AC source range to the AUTO mode. While in AUTO mode, the AC
source automatically switches to the optimal voltage range for the voltage programmed.
Note that switching voltage ranges causes the output to be interrupted momentarily. If
power to the unit under test cannot be interrupted, the AUTO range should not be used.
To determine the actual voltage range in which the unit is operating, the VOLT:RANG?
command query can be used.
For 2001RP-AV models configured for single voltage range, only one range is available. In
this case, the VOLT:RANGe:AUTO command has no meaning and should not be used.
The maximum current limit that can be programmed at a given time depends on the voltage
range setting in which the unit is presently operating. Programming a current limit value that
exceeds the high voltage range capability does NOT cause the voltage range to change,
even while in AUTO mode. Only voltage changes can prompt an automatic voltage range
change.
Command Syntax
[SOURce:]VOLTage:RANGe:AUTO <Bool>
Parameters
<Bool> 0 | 1
*RST Value
0
Examples VOLT:RANG:AUTO
1
Query Syntax
VOLT:RANG:AUTO?
Returned Parameters
<Bool>
Related Commands
VOLT LIM:VOLT? VOLT:RANG:LEV
2001RP April
2007
93
Summary of Contents for 2001RP
Page 33: ...User and Programming Manual Rev N Figure 3 7 Test Setup 2001RP 24 April 2007 2001RP...
Page 47: ......
Page 60: ...User and Programming Manual Rev N Figure 6 3 Internal Adjustments 2001RP April 2007 51...
Page 65: ...User and Programming Manual Rev N Figure 7 1 Assembly Location 56 April 2007 2001RP...
Page 68: ...User and Programming Manual Rev N Figure 7 2 Replaceable Fuse Locations 2001RP April 2007 59...
Page 70: ...User and Programming Manual Rev N Figure 8 1 Rear Panel View 2001RP April 2007 61...
Page 90: ......