4.4 Current Limit Modes
The 2001RP supports two modes of current limiting. In either mode, the end user can set
the RMS current level at which the current limit function will kick in. The available current
limit modes are:
Constant Current (CC) mode
This is the default mode as supplied from the
factory. In this mode of operation, the output
voltage is automatically reduced by the AC power
source to maintain the set current. Once the
current drawn drops below the set limit value, the
voltage is increased again. No error message is
generated in this mode of operation. This mode
is suitable for starting up loads that require high
inrush currents but don’t need full nominal
voltage to run. (example AC motors).
Constant Voltage (CV) mode
In this mode of operation, the output of the AC
source is turned off and the programmed voltage
is set to 0 volt as soon as the load current
exceeds the set current limit value. (Trip current).
The selected current limit mode may be changed by moving a jumper on the AC power
source controller board. This requires that the AC source is disconnected from the AC line
and the top cover is removed. The current mode jumper is located at the top of the
controller board which is directly behind the front panel.
MODE JUMPERS
CC
W1 and W2 Removed
CV
W1 and W2 Installed
2001RP April
2007
35
Summary of Contents for 2001RP
Page 33: ...User and Programming Manual Rev N Figure 3 7 Test Setup 2001RP 24 April 2007 2001RP...
Page 47: ......
Page 60: ...User and Programming Manual Rev N Figure 6 3 Internal Adjustments 2001RP April 2007 51...
Page 65: ...User and Programming Manual Rev N Figure 7 1 Assembly Location 56 April 2007 2001RP...
Page 68: ...User and Programming Manual Rev N Figure 7 2 Replaceable Fuse Locations 2001RP April 2007 59...
Page 70: ...User and Programming Manual Rev N Figure 8 1 Rear Panel View 2001RP April 2007 61...
Page 90: ......