4.3 Setting the Power on Initialization Values
The 2001RP is supplied with default factory settings when the unit is powered up. The
factory settings are:
Parameter
Factory default setting
Voltage range
Low
Auto Voltage Range
OFF
Voltage 5.0
Volt
Frequency
60 Hz ( 400 Hz on 2001RP-AV models)
Current limit
13.4 amps (17.4 amps on 2001RP-AV models, 14.8
amps on 2001RP-LZ models).
Display mode
Frequency
Output Relay state
OFF
Output start state
OFF
Output start phase
0 degrees
Local / Remote State
Local. Front panel unlocked.
Table 4-1: Factory Default Power on Settings
It is possible to change the power on initialization values in one of two ways:
1. Using the optional RS232 or IEEE-488 interface and the supplied PGUI32 program.
2. Using the front panel.
To change the power on initialization values from the front panel, proceed as follows:
1. Set the unit up in the desired way from the front. (Range, voltage, frequency, current
limit, output relay state, display mode).
2. Press and hold the SELECT key (normally toggles LCD display mode). Make sure the
correct indicator LED is illuminated, e.g., FREQ to display voltage and frequency.
3. While holding the SELECT key, press the OUTPUT ON/OFF key. This will save the
present front panel settings in non volatile memory register (NVM) no 7 and assign this
register as the power on register.
4. Release both keys. Note that as step 2 is executed, the display mode will toggle . This
means the display mode after power up is different from how the unit was set up. To
avoid this, just toggle the display till one position before the desired power on mode first
and then execute steps 2-4.
5. This procedure can be repeated as often as needed by the user.
34 April
2007
2001RP
Summary of Contents for 2001RP
Page 33: ...User and Programming Manual Rev N Figure 3 7 Test Setup 2001RP 24 April 2007 2001RP...
Page 47: ......
Page 60: ...User and Programming Manual Rev N Figure 6 3 Internal Adjustments 2001RP April 2007 51...
Page 65: ...User and Programming Manual Rev N Figure 7 1 Assembly Location 56 April 2007 2001RP...
Page 68: ...User and Programming Manual Rev N Figure 7 2 Replaceable Fuse Locations 2001RP April 2007 59...
Page 70: ...User and Programming Manual Rev N Figure 8 1 Rear Panel View 2001RP April 2007 61...
Page 90: ......