User and Programming Manual - Rev N
6.3.3 Measurement Calibration - Full Scale
Note:
Measurement calibration can be done at any output frequency setting
as only one set of calibration coefficients is used for measurement
calibration. However, for best results, calibrate the measurements at
the frequency which is used in most applications, e.g. 50 Hz, 60 Hz
(except model 2001RP-AV) or 400 Hz.
Refer to Table 6-1 for correct load resistors based on model number.
Full Scale Voltage Measurement Calibration
1. Connect the test equipment to the power source as shown in Figure 6-1.
Do not
connect any load during this step. Voltage calibration must be done under no load
conditions.
2. Select the High Voltage Range and program the output voltage to 240.0 volts AC or the
maximum voltage available if less than 240 Volts.
3. Put the 2001RP in CAL mode by pressing the recessed CAL button.
4. Use the SELECT key to select the Voltage Measurement calibration mode. This mode is
indicated by the FREQ LED.
5. Use the left control knob to adjust the voltage reading shown on the right hand LCD until
it reads as close as possible to the reference reading as displayed on the external DMM.
Full Scale Current Measurement Calibration
1. Connect the test equipment to the power source as shown in Figure 6-2.
2. Apply the 11.3
Ω
resistive load to the output terminals. Make sure the external DMM
used for voltage measurement is across the sense terminals.
3. Program the output voltage to a value that produces between 3 and 3.9 amps. Do not
load the output for more than 3.9 amps.
4. Use the SELECT key to select the Current Measurement calibration mode. This mode is
indicated by the I RMS LED.
5. Use the left control knob to adjust the current reading shown on the right hand LCD until
it reads as close as possible to the reference reading indicated by the DMM measuring
the current shunt voltage (DMM 1). Note that .39 V reading represents 3.9 Amps if the
recommended 100 m
Ω
shunt is used.
Full Scale Peak Current Measurement Calibration
1. Connect the test equipment to the power source as shown in Figure 6-2.
2. Select the low voltage range and program the output voltage to a value that produces
about 13 Amps rms.
48 April
2007
2001RP
Summary of Contents for 2001RP
Page 33: ...User and Programming Manual Rev N Figure 3 7 Test Setup 2001RP 24 April 2007 2001RP...
Page 47: ......
Page 60: ...User and Programming Manual Rev N Figure 6 3 Internal Adjustments 2001RP April 2007 51...
Page 65: ...User and Programming Manual Rev N Figure 7 1 Assembly Location 56 April 2007 2001RP...
Page 68: ...User and Programming Manual Rev N Figure 7 2 Replaceable Fuse Locations 2001RP April 2007 59...
Page 70: ...User and Programming Manual Rev N Figure 8 1 Rear Panel View 2001RP April 2007 61...
Page 90: ......