User and Programming Manual - Rev N
3.6 Output
Connections
The output terminal block is located at the rear of the power source or on the front panel with
Option -SKT. All load connections must be made on the rear panel terminal block or from
the front panel socket if the -SKT Option is installed. The remote sense inputs are normally
jumpered directly to the adjacent power output pins on the output connector. If the sense
terminals are not connected to the power output pins, the output voltage may rise
significantly above the programmed value. When connecting the output wires, use the
supplied safety cover/strain relief and mount over the output connector. Also mount the
supplied snap-on ferrite filter around the output wiring between the output terminals and the
strain relief grommet. Enclose ferrite filter with the safety cover. See Figure 3-6.
If the load is a significant distance from the output terminal strip it may be desirable to sense
the output voltage directly at the load. If remote sensing is to be used, then the size of the
output power wiring must be heavy enough to prevent a voltage difference greater than 7%
of the programmed value from the power output pins to the sense input pins. Table 3-1
shows the maximum length of the output wires. The table assumes the Remote Sense input
is connected at the load. For 2001RP models with lower voltage ranges than 150 V
RMS
, the
wire gauge may have to be reduced. Numbers shown in Table 3-1 are for standard 2001RP
model with 150 V
RMS
range.
Maximum
output current
Wire gauge
(AWG)
Conductor nom. diameter
Inches mm
Maximum wire length between output
terminals and load
feet meter
13.4 A
RMS
14
.072 inch 1.93 mm
46 ft. 14 m.
13.4 A
RMS
12
.090 inch 2.27 mm
73 ft. 22 m.
13.4 A
RMS
10
.112 inch 2.84 mm
117ft. 35 m.
Table 3-1: Maximum Output Wiring Lengths
The Remote Sense inputs must be connected or an output voltage 7% higher than the
programmed output will be generated. If the 2001RP is configured for constant voltage, a
fault will be generated.
20 April
2007
2001RP
Summary of Contents for 2001RP
Page 33: ...User and Programming Manual Rev N Figure 3 7 Test Setup 2001RP 24 April 2007 2001RP...
Page 47: ......
Page 60: ...User and Programming Manual Rev N Figure 6 3 Internal Adjustments 2001RP April 2007 51...
Page 65: ...User and Programming Manual Rev N Figure 7 1 Assembly Location 56 April 2007 2001RP...
Page 68: ...User and Programming Manual Rev N Figure 7 2 Replaceable Fuse Locations 2001RP April 2007 59...
Page 70: ...User and Programming Manual Rev N Figure 8 1 Rear Panel View 2001RP April 2007 61...
Page 90: ......