10.3 Measurement Subsystem
This subsystem programs the measurement capability of the RP Series AC source.
Subsystem Syntax
:MEASure
:VOLTage?
Returns the rms value of voltage
:CURRent?
Returns
the
rms
value
of
current
:AMPLitude
:MAX?
Returns
peak
current
:RESet
Reset
the
peak
current
measurements
:CREStfactor?
Returns
current
crest
factor
:RANGe
Selects measurement range 0 = High, 1 = Low, 2 = AUTO
:POWer
[:REAL]? Returns
real
power
in
Watt
:APParent?
Returns
apparent
power
in
VA
PFACtor?
Returns
power
factor
MEASure:VOLTage?
This query returns the AC rms voltage being sourced at the output terminals. Note that the
output relay must be closed or no voltage will be present at the output terminals.
Query Syntax
MEASure:VOLTage?
Parameters None
Examples MEAS:VOLT?
Returned Parameters
<NR2>
Related Commands
OUTP 0
⎜
1
MEASure:CURRent?
This query returns the rms value of the output AC current being sourced at the output
terminals. Note that the output relay must be closed to obtain current flow.
Query Syntax
MEASure:CURRent?
Parameters None
Examples MEAS:CURR?
Returned Parameters
<NR2>
Related Commands
OUTP 0
⎜
1
MEAS:CURR:AMPL:MAX?
2001RP April
2007
87
Summary of Contents for 2001RP
Page 33: ...User and Programming Manual Rev N Figure 3 7 Test Setup 2001RP 24 April 2007 2001RP...
Page 47: ......
Page 60: ...User and Programming Manual Rev N Figure 6 3 Internal Adjustments 2001RP April 2007 51...
Page 65: ...User and Programming Manual Rev N Figure 7 1 Assembly Location 56 April 2007 2001RP...
Page 68: ...User and Programming Manual Rev N Figure 7 2 Replaceable Fuse Locations 2001RP April 2007 59...
Page 70: ...User and Programming Manual Rev N Figure 8 1 Rear Panel View 2001RP April 2007 61...
Page 90: ......