User and Programming Manual - Rev N
2. Program the output voltage to 150V (135V for -LZ option, 115 V for –AV option) on the
low voltage range and 60 Hz (400 Hz for –AV option).
3. Use
the
SELECT
key to select the Current Measurement display. Note that this is not
the Current Measurement calibration mode. Only the I RMS
LED
should be illuminated.
4. Observe the actual output current as shown on the external DMM and current shunt.
Adjust R3 on the Range/Relay Board for a measurement display equal to the actual
output current.
CAUTION:
The amplifier wind tunnel must be covered for proper cooling if the
adjustment takes more than 1 minute.
6.4.2 Current Limit Adjustment
Note:
If the Current Limit and High Current Measurement Range
Adjustments are both adjusted, the High Current Measurement Range
must be adjusted first.
1. Connect the test equipment as shown in Figure 6-2. Use the 11.3 ohm load.
2. Program the current to 10.0 A. Program the output to 150V (135V for the -LZ option,
115 V for the –AV option) on the low voltage range and 60 Hz. (400 Hz for the –AV
option).
CAUTION:
The amplifier wind tunnel must be covered for proper cooling if the
adjustment takes more than 1 minute.
1. Adjust R50 on the Oscillator Board for an output current of 10.5 A.
50 April
2007
2001RP
Summary of Contents for 2001RP
Page 33: ...User and Programming Manual Rev N Figure 3 7 Test Setup 2001RP 24 April 2007 2001RP...
Page 47: ......
Page 60: ...User and Programming Manual Rev N Figure 6 3 Internal Adjustments 2001RP April 2007 51...
Page 65: ...User and Programming Manual Rev N Figure 7 1 Assembly Location 56 April 2007 2001RP...
Page 68: ...User and Programming Manual Rev N Figure 7 2 Replaceable Fuse Locations 2001RP April 2007 59...
Page 70: ...User and Programming Manual Rev N Figure 8 1 Rear Panel View 2001RP April 2007 61...
Page 90: ......