CALIBRATION PROCEDURE

PXIe-4135

Single-Channel System Source-Measure Unit (SMU)

This document contains the verification and adjustment procedures for the PXIe-4135. Refer
to 

ni.com/calibration

 for more information about calibration solutions.

Please review and become familiar with the entire procedure before beginning the calibration
process.

Contents

Required Software.....................................................................................................................2
Related Documentation.............................................................................................................2
Password................................................................................................................................... 2
Calibration Interval................................................................................................................... 3
Test Equipment..........................................................................................................................3
Test Conditions..........................................................................................................................4
Safety Guidelines for System Operation...................................................................................5
As-Found and As-Left Limits................................................................................................... 5
Calibration Overview................................................................................................................6
Verification................................................................................................................................6

Self-Calibrating the PXIe-4135........................................................................................ 7
Testing the Safety Interlock.............................................................................................. 7
Connecting and Configuring Equipment for Voltage Verification....................................7
Verifying Voltage Measurement and Output.....................................................................8
Verifying Remote Sense Voltage Offset............................................................................9
Verifying Voltage Remote Sense.....................................................................................10
Verifying Current Offset Null......................................................................................... 12
Verifying Current Offset................................................................................................. 13
Verifying Load Regulation..............................................................................................14
Verifying 10 nA Current Measurement and Output........................................................16
Verifying 1 μA Current Measurement and Output..........................................................17
Verifying 100 μA to 100 mA Current Measurement and Output....................................19
Verifying 1 A and 3 A Current Measurement and Output.............................................. 21

Adjustment.............................................................................................................................. 23

Adjusted Specifications...................................................................................................23
Initiating the Adjustment Session................................................................................... 23
Voltage and Current Output............................................................................................ 23
Residual Offset Voltage...................................................................................................27
Closing the Adjustment Session..................................................................................... 28
Alternative to Performing Adjustment Procedures.........................................................28

Summary of Contents for PXIe-4135

Page 1: ...Self Calibrating the PXIe 4135 7 Testing the Safety Interlock 7 Connecting and Configuring Equipment for Voltage Verification 7 Verifying Voltage Measurement and Output 8 Verifying Remote Sense Volta...

Page 2: ...ation support in the locations shown in the following table ADE Calibration Support Location LabVIEW NI DCPower Calibration palette LabWindows CVI NI DCPower function panel niDCPower fp You can downlo...

Page 3: ...Offset Accuracy Current Measurement and Output Accuracy Voltage 9 ppm accuracy and 100 nV resolution Current 25 ppm accuracy and 10 pA resolution 100 M current shunt IET Labs SRL 100M Pom5219 10 nA c...

Page 4: ...x cabling for all HI and HI Sense connections For LO and LO Sense connections use shielded twisted pair copper wire for all cable connections to the device To ensure the system has had adequate time t...

Page 5: ...s Ensure proper warnings and signage exist for workers in the area of operation Provide training to all system operators so that they understand the potential hazards and how to protect themselves Ins...

Page 6: ...mation Explorer MAX 2 Verification Verify the existing operation of the PXIe 4135 This step confirms whether the PXIe 4135 is operating within the published specifications prior to adjustment 3 Adjust...

Page 7: ...connector from the PXIe 4135 front panel 2 Ensure that the safety interlock input on the test fixture is closed 3 Set the niDCPower Output Function property or NIDCPOWER_OUTPUT_FUNCTION attribute to D...

Page 8: ...ompare a set of voltages measured by a DMM to the voltage test points requested by the PXIe 4135 Verify level ranges in the specified order Refer to the following table as you complete the following s...

Page 9: ...d test point 5 Compare a DMM voltage measurement to the voltage measurement test limits a Take a voltage measurement using the DMM b Calculate the lower and upper voltage measurement test limits using...

Page 10: ...specified repeat the previous steps for each test point from setting the level to the test point on the PXIe 4135 up to this step 6 If more than one level range is specified repeat the previous steps...

Page 11: ...by more than 1 C in the previous 5 minutes b After the internal temperature has stabilized if the temperature still exceeds Tcal 1 C call the self calibration VI or function 6 Set the level on the PX...

Page 12: ...eft Offset Test Limit 600 mV 10 nA 10 nA 0 mV 750 fA3 1 Disconnect all equipment from the output of the PXIe 4135 2 Measure the internal device temperature and perform self calibration if necessary a...

Page 13: ...00 mV 10 nA 10 nA 0 mV 5 pA 1 pA 1 A 1 A 40 pA 17 pA 100 A 100 A 2 nA 970 pA 1 mA 1 mA 20 nA 9 7 nA 10 mA 10 mA 200 nA 97 nA 100 mA 100 mA 2 A 970 nA 1 A 1 A 20 A 8 5 A 3 A 1 A 600 A 85 A 200 V 10 nA...

Page 14: ...an authorized NI service representative to request a Return Material Authorization RMA Refer to the following table as you complete the following steps Table 7 Load Regulation Verification Level Range...

Page 15: ...After measuring the internal device temperature wait for the internal device temperature to stabilize Temperature is considered stable when it has not changed by more than 1 C in the previous 5 minute...

Page 16: ...the following figure Figure 6 Current Connection Diagram Part 1 Triax Cable Triax to Bare Wire Precision Shunt LO LOS HI Triax HIS Triax NI DCPower Device Sense LO Sense HI Triax HI Triax LO CHS GND...

Page 17: ...tput on Figure 7 Current Connection Diagram Part 2 Triax Cable Precision Shunt LO LOS HI Triax HIS Triax NI DCPower Device Sense LO Sense HI Triax HI Triax LO CHS GND Shielded Combicon to Banana Bare...

Page 18: ...and perform self calibration if necessary a After measuring the internal device temperature wait for the internal device temperature to stabilize Temperature is considered stable when it has not chang...

Page 19: ...steps 1 11 for each test point in the previous table Verifying 100 A to 100 mA Current Measurement and Output Compare a set of currents measured by a DMM to the current test points requested by the PX...

Page 20: ...er Output Function property or NIDCPOWER_OUTPUT_FUNCTION attribute to DC Current for the PXIe 4135 3 Set the first specified level range limit range and limit on the PXIe 4135 4 Measure the internal d...

Page 21: ...A Current Measurement and Output Compare a set of currents measured by a DMM to the current test points requested by the PXIe 4135 Complete this procedure only after successfully completing all previ...

Page 22: ...o the first specified test point 6 Complete the following 4 steps within 5 minutes or less of completing the previous step in order to ensure the internal device temperature remains stable 7 Calculate...

Page 23: ...mum of five minutes for the internal device temperature to stabilize 2 Initiate an external calibration session a specific type of NI DCPower session by calling the niDCPower Initialize External Calib...

Page 24: ...t Point 6 V 100 mA 5 V 5 V 1 Set the first specified level range limit range and limit on the PXIe 4135 2 Set the level on the PXIe 4135 to the first specified test point 3 Take a voltage measurement...

Page 25: ...iax LO CHS GND 2 Set the niDCPower Output Function property or NIDCPOWER_OUTPUT_FUNCTION attribute to DC Current for the PXIe 4135 3 Set the first specified level range limit range and limit on the PX...

Page 26: ...te this procedure only after successfully completing all previous adjustment procedures Adjust ranges in the specified order Refer to the following table as you complete the following steps Table 15 1...

Page 27: ...calling the niDCPower Cal Adjust Current Limit VI or niDCPower_CalAdjustCurrentLimit function a Input the calculated shunt current measurements as the measured outputs b Input the test points as the r...

Page 28: ...rval 1 Call either the niDCPower Initialize External Calibration VI or the niDCPower_InitExtCal function 2 Call either the niDCPower Close External Calibration VI or the niDCPower_CloseExtCal function...

Page 29: ...Offset Null task Verifying Current Offset Updated as left offset test limits and added instructions to reset PLCs NI Services Visit ni com support to find support resources including documentation dow...

Page 30: ...s and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy and how to...