Detailed Device Description
Physician Manual
en
Stimulator Programmable Characteristics
The stimulation parameters are independent for the two DBS Leads such that
stimulation of two different brain targets can have different amplitudes, pulse widths,
stimulation rates, and contact configurations. The two DBS Leads can also have
differing monopolar and multipolar configurations; however, a given DBS Lead
contact cannot be programmed as both monopolar and multipolar.
The programmable parameter ranges for the Stimulator are shown in Table 2.
Note:
Some frequency combinations may not be used. See the DBS Programming
Manual for specific frequency combinations.
Table 2: Stimulator Function
Parameter
Range
Amplitude
1
0.1 – 20 mA
Rate
2, 3
2 – 255 Hz
Pulse Width
4
10 – 450 µs
Cycle (On/Off)
1 s – 90 min
Stim Ramp On
1 – 10 s
Contact Connections
16
Independent Areas of Stim (4 Programs with
4 Areas per Program)
16
1
The programmable coverage for each individual contact is limited to 12.7 mA.
A programming interlock is enforced to limit the total output current to 20 mA or less
per coverage area. For example, a maximum current output of 12.7 mA on one contact
would limit the total summed current output on the remaining contacts to 7.3 mA within
one coverage area.
2
The rate is limited to 255 Hz for a given area.
3
The use of higher frequencies than the ones that are established (250 Hz) is the sole
responsibility of the user.
4
The use of lower pulse widths than established (60-450 μsec) is the sole responsibility
of the user.
NOTE:
While it is possible to program stimulation rates higher than 255 Hz,
clinical effectiveness and saftey of these cumulative rates has not been
demonstrated and is not recommended.
WARNING:
Although the CP utilizes warnings to aid clinicians in their
determination of safe charge density levels, the primary responsibility
of keeping safe current levels resides with the clinician. Additionally,
maximum and minimum amplitude levels allowed by the Remote
Control should be set and verified by the physician to ensure that these
levels are safe.
Summary of Contents for DB-1110-C
Page 345: ...Anhang Handbuch Arzt 92215483 02 341 von 1717 de Diese Seite wurde absichtlich frei gehalten ...
Page 769: ...Tillegg Legens håndbok 92215483 02 765 av 1717 no Denne siden er blank med hensikt ...
Page 855: ...Appendiks Lægehåndbog 92215483 02 851 af 1717 da Denne side er tilsigtet tom ...
Page 1109: ...Teknik Destek Hekim Kılavuzu 92215483 02 1105 1717 tr Bu sayfa özellikle boş bırakılmıştır ...
Page 1197: ... Руководство для врача 92215483 02 1193 из 1717 ru Эта страница намеренно оставлена пустой ...
Page 1635: ...Dodatek Podręcznik lekarza 92215483 02 1631 z 1717 pl Strona celowo pozostawiona pusta ...
Page 1721: ... en Physician Manual 92215483 02 1717 of 1717 This Page Intentionally Left Blank ...