Stimulator Implantation
Physician Manual
en
CAUTION:
Verify proper DBS Extension insertion by checking impedances before
tightening the Setscrew. Tightening the Setscrew down onto a contact
can damage the DBS Extension.
2. Check impedances to verify connections before tightening the Setscrew.
a. Place the Stimulator partially in the subcutaneous pocket.
b. Test impedances using the Remote Control or Clinician Programmer.
3. Verify that the retention sleeve on the DBS Extension is located directly under the
Setscrew in the Stimulator header.
Correct
Incorrect
Note:
The retention sleeve is easily distinguishable from the contacts by its length
(see below).
Contacts Retention
Sleeve
Proximal End of the DBS Lead
4. Pass the Torque Wrench through the slit in the septum located on the top of the
Stimulator header.
5. Tighten the Setscrew in the Stimulator until the Torque Wrench clicks, indicating
the Setscrew is fully secured.
Note:
To tighten the Setscrew, rotate the Torque Wrench clockwise. To loosen the
Setscrew, rotate the Torque Wrench counterclockwise.
Summary of Contents for DB-1110-C
Page 345: ...Anhang Handbuch Arzt 92215483 02 341 von 1717 de Diese Seite wurde absichtlich frei gehalten ...
Page 769: ...Tillegg Legens håndbok 92215483 02 765 av 1717 no Denne siden er blank med hensikt ...
Page 855: ...Appendiks Lægehåndbog 92215483 02 851 af 1717 da Denne side er tilsigtet tom ...
Page 1109: ...Teknik Destek Hekim Kılavuzu 92215483 02 1105 1717 tr Bu sayfa özellikle boş bırakılmıştır ...
Page 1197: ... Руководство для врача 92215483 02 1193 из 1717 ru Эта страница намеренно оставлена пустой ...
Page 1635: ...Dodatek Podręcznik lekarza 92215483 02 1631 z 1717 pl Strona celowo pozostawiona pusta ...
Page 1721: ... en Physician Manual 92215483 02 1717 of 1717 This Page Intentionally Left Blank ...