Physician Manual
Physician Manual
92215483-02 4 of 1717
en
Note:
Vercise DBS Lead-only system (before Stimulator is implanted) is MR
conditional. An MRI examination can be conducted safely when all
instructions in the supplemental manual MRI Guidelines for Boston Scientific
DBS Systems are followed.
For the latest version of the manual go to www.bostonscientific.com/manuals.
Electromagnetic Interference.
Strong electromagnetic fields can potentially turn
the Stimulator off, cause temporary unpredictable changes in stimulation, or interfere
with the Remote Control communication. Patients should be counseled to avoid or
exercise care around:
•
Theft detectors such as those used at entrances/exits of department stores,
libraries, and other public establishments. The patient should proceed with
caution, ensuring to move through the center of the detector as quickly as
possible.
•
Security screeners, such as those used in Airport Security or at entrances
to government buildings, including hand-held scanners. The patient should
request assistance to bypass the device. If the patient must pass through the
security screener, they should move quickly through the device staying as far
from the physical device as allowable.
•
Power lines or power generators.
•
Electric steel furnaces and arc welders.
•
Large magnetized stereo speakers.
•
Strong magnets.
•
Automobiles or other motorized vehicles using a LoJack system or other
anti-theft systems that can broadcast a radio frequency (RF) signal. The high
energy fields produced by these systems may interfere with the operation of
the Remote Control and its ability to control stimulation.
•
Other sources of electromagnetic interference, such as RF transmitters at
television or radio broadcast stations, Amateur Radio or Citizens Band radio
transceivers, or Family Radio Service band transceivers.
DBS Extension Connector and Stimulator Placement.
Implanting the DBS
Extension connector in the soft tissue of the neck may increase the chance of DBS
Lead breakage. Boston Scientific recommends placing the DBS Extension connector
behind the ear such that glasses or headgear do not interfere with the system.
Boston Scientific recommends that the Stimulator be placed subclavicularly.
Heat Due to Charging.
The Charger may become warm while charging the
Stimulator. The Charger should be handled with care. Failure to use either the
Charging Collar or an Adhesive Patch while charging, as directed, may result in
a burn. The Patient should not charge while sleeping. This may result in a burn. If the
patient experiences pain or discomfort, they should cease charging and contact their
physician.
Summary of Contents for DB-1110-C
Page 345: ...Anhang Handbuch Arzt 92215483 02 341 von 1717 de Diese Seite wurde absichtlich frei gehalten ...
Page 769: ...Tillegg Legens håndbok 92215483 02 765 av 1717 no Denne siden er blank med hensikt ...
Page 855: ...Appendiks Lægehåndbog 92215483 02 851 af 1717 da Denne side er tilsigtet tom ...
Page 1109: ...Teknik Destek Hekim Kılavuzu 92215483 02 1105 1717 tr Bu sayfa özellikle boş bırakılmıştır ...
Page 1197: ... Руководство для врача 92215483 02 1193 из 1717 ru Эта страница намеренно оставлена пустой ...
Page 1635: ...Dodatek Podręcznik lekarza 92215483 02 1631 z 1717 pl Strona celowo pozostawiona pusta ...
Page 1721: ... en Physician Manual 92215483 02 1717 of 1717 This Page Intentionally Left Blank ...