DBS Lead Implantation
Physician Manual
en
c. Remove the stereotactic frame.
d. Remove the stylet.
WARNING:
Do not reinsert the stylet into the DBS Lead while the DBS Lead is
in the brain, as this may damage the DBS Lead and/or cause patient
harm.
6. Verify that the DBS Lead has not moved from the desired location.
7. Place the split suture sleeve onto the DBS Lead. Be careful not to move the DBS
Lead when placing the suture sleeve.
8. Rotate the suture sleeve until the slit is facing the skull.
Note:
If the slit is not located properly, the DBS Lead may not be protected by the
suture sleeve and may be in direct contact with the mini plate.
9. Place a mini plate (not provided) over the suture sleeve and DBS Lead.
CAUTION:
Do not place the mini plate in direct contact with the DBS Lead.
Ensure that the suture sleeve lies between the mini plate and the
DBS Lead.
10. Secure the mini plate to the skull per the manufacturer’s instructions.
Summary of Contents for DB-1110-C
Page 345: ...Anhang Handbuch Arzt 92215483 02 341 von 1717 de Diese Seite wurde absichtlich frei gehalten ...
Page 769: ...Tillegg Legens håndbok 92215483 02 765 av 1717 no Denne siden er blank med hensikt ...
Page 855: ...Appendiks Lægehåndbog 92215483 02 851 af 1717 da Denne side er tilsigtet tom ...
Page 1109: ...Teknik Destek Hekim Kılavuzu 92215483 02 1105 1717 tr Bu sayfa özellikle boş bırakılmıştır ...
Page 1197: ... Руководство для врача 92215483 02 1193 из 1717 ru Эта страница намеренно оставлена пустой ...
Page 1635: ...Dodatek Podręcznik lekarza 92215483 02 1631 z 1717 pl Strona celowo pozostawiona pusta ...
Page 1721: ... en Physician Manual 92215483 02 1717 of 1717 This Page Intentionally Left Blank ...