Medical Devices/Therapies.
The following medical therapies or procedures may
turn stimulation off, cause permanent damage to the Stimulator, or may cause injury
to the patient:
•
Electrocautery – Electrocautery can transfer destructive current into the
DBS Leads and/or Stimulator.
•
External Defibrillation – Safe usage of external defibrillation has not been
established and damage should be ascertained following defibrillation.
•
Lithotripsy – High frequency signals directed near the Stimulator may damage
circuitry.
•
Radiation Therapy – Lead shielding should be used over the Stimulator to
prevent damage from high radiation. Any damage to the device by radiation
may not be immediately detectable.
•
Transcranial Stimulation – Safe use of electromagnetic therapies, such as
transcranial magnetic stimulation, have not been established.
•
MRI – Patients implanted with the full Vercise DBS System (leads, extensions
and stimulator) should not be subjected to MRI to avoid damage to the device
and patient harm.
Note:
Vercise DBS Lead-only system (before Stimulator is implanted) is MR
conditional. An MRI examination can be conducted safely when all
instructions in the supplemental manual MRI Guidelines for Boston Scientific
DBS Systems are followed.
For the latest version of the manual go to www.bostonscientific.com/manuals.
•
Diathermy – The energy generated by diathermy can be transferred to the
Vercise™ DBS System and may result of damage to the device or patient
harm.
If any of the above is required by medical necessity, the procedure(s) should be
performed as far from the implanted components as possible. Ultimately, however,
the Stimulator may require explantation as a result of damage to the device or patient
harm.
Sterilization.
Contents of the surgical kits are supplied sterile using an ethylene
oxide process. Do not use if sterile barrier is damaged. If damage is found, call your
Boston Scientific representative and return the damaged part to Boston Scientific.
Resterilization.
For single patient use only. Do not reuse, reprocess or
resterilize. Reuse, reprocessing or resterilization may compromise the structural
integrity of the device and/or lead to device failure which, in turn, may result in
patient injury, illness or death. Reuse, reprocessing or resterilization may also create
a risk of contamination of the device and/or cause patient infection or cross-infection,
including, but not limited to, the transmission of infectious disease(s) from one
patient to another. Contamination of the device may lead to injury, illness or death of
the patient. If resterilization is desired, obtain a new sterile device for implantation.
Summary of Contents for DB-1110-C
Page 345: ...Anhang Handbuch Arzt 92215483 02 341 von 1717 de Diese Seite wurde absichtlich frei gehalten ...
Page 769: ...Tillegg Legens håndbok 92215483 02 765 av 1717 no Denne siden er blank med hensikt ...
Page 855: ...Appendiks Lægehåndbog 92215483 02 851 af 1717 da Denne side er tilsigtet tom ...
Page 1109: ...Teknik Destek Hekim Kılavuzu 92215483 02 1105 1717 tr Bu sayfa özellikle boş bırakılmıştır ...
Page 1197: ... Руководство для врача 92215483 02 1193 из 1717 ru Эта страница намеренно оставлена пустой ...
Page 1635: ...Dodatek Podręcznik lekarza 92215483 02 1631 z 1717 pl Strona celowo pozostawiona pusta ...
Page 1721: ... en Physician Manual 92215483 02 1717 of 1717 This Page Intentionally Left Blank ...