Detailed Device Description
Physician Manual
en
Programmer Communication
The Clinician Programmer (CP) can communicate with a Stimulator. In order to
begin a programming session, the CP and the Remote Control Infrared (IR) windows
must be aligned.
1. Turn on the CP and the Bionic Navigator™ Software will boot-up.
2. Enter the Clinician Options menu on the Remote Control by pressing the
and
P
buttons simultaneously.
3. Select CP Mode.
4. When the Remote Control displays “CP Ready,”
put the IR Interface and Remote Control into the IR
Interface Holder to keep the IR windows aligned.
Note:
Orient the IR Interface with the yellow circle
facing down to maximize communication.
5. Plug the serial end of the USB to Serial Adapter into the
IR Interface.
6. Plug the USB to Serial Adapter into the USB port on
the CP.
7. Position the patient within 60 cm (2 ft) of the Remote
Control to ensure a complete communication link
from the programmer to the Stimulator (see
Remote
Control Position
Note:
A message will be displayed on the Clinician
Programmer if the communication range is not
satisfactory.
For more detailed instructions on how to use the Clinician Programmer and Bionic
Navigator Software with the Vercise™ DBS System, see the DBS Programming
Manual.
Summary of Contents for DB-1110-C
Page 345: ...Anhang Handbuch Arzt 92215483 02 341 von 1717 de Diese Seite wurde absichtlich frei gehalten ...
Page 769: ...Tillegg Legens håndbok 92215483 02 765 av 1717 no Denne siden er blank med hensikt ...
Page 855: ...Appendiks Lægehåndbog 92215483 02 851 af 1717 da Denne side er tilsigtet tom ...
Page 1109: ...Teknik Destek Hekim Kılavuzu 92215483 02 1105 1717 tr Bu sayfa özellikle boş bırakılmıştır ...
Page 1197: ... Руководство для врача 92215483 02 1193 из 1717 ru Эта страница намеренно оставлена пустой ...
Page 1635: ...Dodatek Podręcznik lekarza 92215483 02 1631 z 1717 pl Strona celowo pozostawiona pusta ...
Page 1721: ... en Physician Manual 92215483 02 1717 of 1717 This Page Intentionally Left Blank ...