D-132
Chapter D - Software
Automated Chemistry Analyzer
AU2700 User's Guide
May, 2002
3.7.1 Protocol
OLU1329E
F4 (Set)
Unit No.:
Set Up
Protocol
Test No.
ALARMCLEAR
?
STANDBY
Help
Exit
Set
Online
1997/09/03 10:10
Error Control
Test Requisition Inf.
Receive:
Results Transfer:
Upper Protocol
Character Format
Character Length:
Parity Bit:
Bit/Sec.:
Stop Bit:
Text Length:
ETB Control
BCC Check
Data Format
Start Code:
End Code:
T4:
T3:
Time Out (x100msec.)
T1:
T2:
T7:
T5:
T6:
Communication Control
Class:
Retry:
Lower Protocol
Data Format:
Zero Suppress:
Rack No.:
Text Format
ALARMCLEAR
Help
Exit
Special
Special
Description:
This screen is used to view or set the online communication protocols. Change the display to
the “Online : Protocol tab” screen by selecting the “Protocol” tab.
Test Requisition Information Receive:
Sets this operation to “Continue” or “Stop”. To change
the setting, select one of them from the pull-down list.
Results Transfer:
Sets this operation to “Continue” or “Stop”. To change the setting, select one
of them from the pull-down list.
Rack No.:
Sets this number to “4”, “5”, or “No”. To change the setting, select one of them
from the pull-down list.
Data Format:
Sets this format to “6” or “9”. To change the setting, select one of them from the
pull-down list.
Zero Suppress:
Sets this function to “Yes” or “No”. To change the setting, select one of them
from the pull-down list.
Character Length:
Sets this length to “7” or “8”. To change the setting, select one of them
from the pull-down list.
Parity Bit:
Sets this bit to “Even”, “Odd”, or “No”. To change the setting, select one of them
from the pull-down list.
Stop Bit:
Sets this bit to “1” or “2”. To change the setting, select one of them from the pull-
down list.
Start Code - left pull-down list:
Sets the first byte of the Start code within the range of “01h”
to “1Fh”. To change the setting, select one of them from the pull-down list.
Start Code - right pull-down list:
Sets the second byte of the Start code to “No” or within the
range of “01h” to “1Fh”. To change the setting, select one of them from the pull-down list.
End Code - left pull-down list:
Sets the first byte of the End code within the range of “01h” to
“1Fh”. To change the setting, select one of them from the pull-down list.
End Code - right pull-down list:
Sets the second byte of the End code to “No” or within the
range of “01h” to “1Fh”. To change the setting, select one of them from the pull-down list.