D-196
Chapter D - Software
Automated Chemistry Analyzer
AU2700 User's Guide
May, 2002
5.5.5 File Management
(M)Maintenance
(M)Maker Maintenance
(F)File Management
ALARMCLEAR
?
STANDBY
Help
Exit
Execute
File Management
11/30/2000 10:10
OLP1file
Operation
Save files to FD
Load files from FD
Files
Test Name
Profile
Round
Test Method
Test Calculate
Test Inter-Check
Test Blank
Common Repeat
Test Repeat
Common QC
Test QC
QC Auto
Calibration Inf
Test Calibration
Cal Auto
Online Inf
Printer Inf
Requisition Inf
Report Format
List Format
Description:
Use this screen to save/load parameters to/from a floppy disk.
Use this software screen to:
Save or load parameters to/from a floppy:
Instruction
If a floppy is removed from the drive while data is being saved, the data being saved or stored in the
floppy may be lost. Floppy disks for parameter storage have an “AU2700 PARAM” volume label.
Initialize a 3.5 floppy disk as a parameters disk one time before parameters are saved or loaded. The
floppy disk can be 2HD or 2DD. Select [Maintenance], [Data Operation], [FD Data Management],
FD Data Initialize Tab. Select the “Initialize Data FD” option. Insert the floppy disk in the computer.
Select function key F5: Start Initialize. From the “Insert FD, Start” prompt, select yes. The message
“Operating: Wait for a while” displays.
1. Select [Maintenance], [Maker Maintenance], [File Management].
2. Under the "Operation" column, select either "Save Files to FD" or "Load Files from FD."
3. Under the "Files" column, select the parameters to save or load.
4. Select function key F5 (execute) and insert the floppy.
Start? Yes.
5. The selected files will revert to the background color when the operation is complete.