Automated Chemistry Analyzer
AU2700 User's Guide
Chapter D - Software
D-187
May, 2002
5.5.2 Analyzer Diagnostics
(M)Maintenance
(M)Maker Maintenance
(A)ANL Diagnostics
ALARMCLEAR
Help
Exit
ANL Diag
OLP1Adiag
Sample Probe Reagent Probe Reagent Refrigerator
Sample Syringe Reagent Syringe
A/Reset
B/Up-Down Step
C/Rotation Step
D/Up-Down Pulse
E/Rotation Pulse
F/Sequence
G/Reset
H/Up-Down
I/Reset Q/Reset
J/Up-Down Step R/Rotation
K/Rotation Step S/Pulse
L/Up-Down Pulse
M/Rotation Pulse
All Pump
N/Sequence
T/On
U/Off
O/Reset
P/Up-Down
Probe/Syringe etc. Mix/Wheel/Nozzle Sample Set Unit 1 Sample Set Unit 2 Combination
Move the cursor and press enter key or click the mouse to select.
?
STANDBY
11/30/2000 10:10
Description
This menu is used by authorized service personnel for service, inspection, and repair work
performed on this system.
Use this software screen to:
Perform analyzer diagnostics:
The analyzer must be in a stop mode before diagnostics can be accessed. The menu consists of 3
pages of mostly hardware components which can be reset or moved in their normal operation to
test proper operation when troubleshooting.
1. Select [Maintenance], [Maker Maintenance], [ANL Diag.].
2. Use your mouse to select the item to check.
Explanations of each function follow this
procedure.
3. Press the STAT ROTATION/DIAG Button on the analyzer to perform the selection.
Explanation of functions:
Sample Probe
A/Reset:
Aligns the sample probe in its home position.
B/Up/Down Step:
Moves the sample probe 1 step vertically (up/down).
C/Rotation Step:
Moves one step between the wash station and items selected. Select the
following: From (rack, cuvette, STAT table, or wash/STD table) and To (cuvette, or ISE).