3.10.2 System
OLP3048
F4
Referring to system
setup parameters
Setting system
setup parameters
(P)Parameters
(S)System
(S)System
One of the tabs
can be selected
Button function
Described in
another section.
Press this key.
Main operation flow
Operation flow
ALARMCLEAR
?
STANDBY
Help
Exit
Set.
System
11/30/2000 10:10
Routine: Sequential Auto/Standard Repeat: Standard
Test Requisition
Barcode
Serum Urine Others
Rack No. Limit
Emergency: Sequential
STAT: Sequential
Type: NW7
Digits:
Check Mode: No (No Chk. Chr)
Routine: 0100 9999
Emergency: 0100 9999
Repeat Routine: 0100 9999
Repeat Emergency:
OLP1sys
Description:
Use this menu to set the analysis mode for routine, emergency and STAT samples. Analysis
mode options:
1. Sequential
2. Rack No. (Rack ID number determines sample number. Example Rack #1 = 1-10, #2 =
11 - 20 etc.)
3. Barcode
The barcode parameters are programmed. The upper rack ID limit number for serum, urine, or
other samples is designated. The repeat mode is set to auto or standard (manual).
Caution
When Test Requisition is set to Sequential, check the setting for Test Requisition Information Receive in
[Parameters] [Online]. If it is set to Real Time, the instrument will query the host computer for the next
sample requisition in the database and test the sample accordingly unless a manual requisition has been
entered. This can result in the wrong S.ID being associated with the test results.
Test requisition:
Set an analysis method. Possible choices are: “Sequential,” “Rack No.,” or
“Barcode.”
See Also
For detailed information about analysis methods, refer to the System Outline Chapter.
Barcode Specification:
Set a type of barcode, the number of digits, and a check mode.
The following restrictions apply if either ISBT-128 or Mixed is selected from the barcode types:
If ISBT-128 is selected:
Check that the mode is set to "Yes"
•