D-192
Chapter D - Software
Automated Chemistry Analyzer
AU2700 User's Guide
May, 2002
5.5.3 ISE Diagnostics
(M)Maintenance
(M)Maker Maintenance
(I)ISE Diagnostics
ALARMCLEAR
Help
Exit
ISE Diag
OLP1idiag
Prime
Calibration Selectivity
Measurement
D. Pot / Electrode
Reset
A/Total Prime
B/Buffer Prime
C/MID/REF Prime
D/Drain-Electrode
E/Drain-Bypass
F/Mix
G/Overflow Sensor
I/Calibration for serum
J/Calibration for urine
K/Cal. for Serum + Urine
L/Selectivity
M/Reset
N/Rack
O/STAT
P/Sequence 1 (Rack)
Q/Sequence 1 (STAT)
R/Sequence 2
S/Sequence 3
Cell: All
Move the cursor and pres Enter key or click the mouse to select.
?
STANDBY
11/30/2000 10:10
Description:
This menu is used by authorized service personnel for service, inspection, and repair work
performed on this system.
Use this software screen to:
Perform ISE Diagnostics:
The analyzer must be in a stop mode before diagnostics can be accessed. The menu consists of
functions that test, reset, or move ISE components for troubleshooting purposes.
1. Select [Maintenance], [Maker Maintenance], [ISE Diag.].
2. Using your mouse, select the item to check.
Explanations of each function follow
procedures.
3. Press the STAT ROTATION/DIAG Button on the analyzer to perform the selection.
Perform a calibration from diagnostics:
1. Place the analyzer in a stop mode. For information on preparing for a calibration, see the
"Performing Calibrations and a Reagent Blank" section located in the Basic Operations
Chapter.
2. Select [Maintenance], [Maker Maintenance], [ISE Diag].
3. The calibration begins as soon as one of the following is selected:
I/Calibration for
serum, J/Calibration for urine, K/Calibration for serum +urine.