Automated Chemistry Analyzer
AU2700 User's Guide
Chapter D - Software
D-113
May, 2002
3.5.3 STAT Table QC
(P) Parameters
(Q) QC Control
(S) STAT Table QC
Position
Test
ALARMCLEAR
?
STANDBY
Help
Exit
Set
STAT Table QC
11/30/2000
10:10
Round Name 1. Chemistry
Operation: Manual
Position
Type
Control
Control Name
No.
No.
OLP1statq
Round No.
21. Serum
1.
Beckman Liquid L1
22. Serum
2.
Beckman Liquid L2
Description:
Use this screen to set the parameters for quality control analysis using the STAT table.
Caution
It is recommended that the operator determine the acceptable length of time to leave samples, calibrators,
and controls on the STAT table. Although the STAT table is cooled, it should not be used to store
samples, calibrators, and controls for an extended period of time. Do not open and close the STAT table
cover frequently, this will increase the temperature.
See Also
For more information, refer to "Quality Control Analysis" and "How to Reprint Reagent Blank,
Calibrator, QC and Patient Data" located in the Basic Operations Chapter. Also, refer to "Quality
Control" located in the Specifications Chapter.
Round Name:
Displays the selected round name. Use the pull-down menu or </> buttons to
select a round.
Operation:
Specifies repeat run QC options for the STAT table. To enable the auto repeat run,
set to "Auto." To enable the manual repeat run, set to "Manual." To disable repeat runs on the
STAT table, set to "None."
Position No., Type, Control No., Control Name:
Displays the position number, type, control
number, and control name for each control.