Automated Chemistry Analyzer
AU2700 User's Guide
Chapter D - Software
D-189
May, 2002
Select the tab located in the upper left corner of the screen called Mix/Wheel/Nozzle.
Mix
A/Reset:
Aligns the mix bar unit in its home position.
B/Up and Down Step:
Moves the mix unit 1 step vertically (up/down).
C/Rotation Step:
Moves the mix unit horizontally to the next position.
D/Up-Down Pulse:
Pulses the mix unit the number of steps specified from 1 to 999 (Up/
Down).
E/Rotation Pulse:
Pulses the mix unit clockwise or counterclockwise the number of steps
specified from 1 to 999.
F/Mix Rotation:
Spins the mix bars.
G/Sequence:
Performs a complete sequence from mixing to moving to the next position.
Wash Nozzle
H/Reset:
Aligns the wash nozzle in its home position.
I/Up and Down Step:
Moves the wash nozzle unit 1 step vertically (up/down).
J/Up-Down Pulse:
Pulses the wash nozzle unit the number of steps specified from 1 to 999
(Up/Down).
K/Sequence:
Performs a wash nozzle sequence.
Cuvette Wheel
O/Reset:
Aligns the cuvette wheel in its home position.
P/Rotation:
Rotates the cuvette wheel according to user specifications. (Clockwise,
counterclockwise, step 1 to 165)
Q/Pulse:
Pulses the cuvette wheel the number of steps specified from 1 to 999.
R/Magnet Off:
Removes power to the pulse motor. To reset select "O/Reset."
Select the tab located in the upper left corner of the screen called Sample Set Unit 1.
Rack Feed
A/Normal:
Feeds rack(s) onto the analyzer in steps. Press the Rev/Diag Button for each
step.
B/Auto Repeat Run: Feeds repeat rack(s) onto the analyzer.
Rack Feeder
C/Reset:
Resets the rack feeder.
D/Forward:
Moves rack(s) forward.
E/Back:
Moves rack(s) backward.
STAT
F/Reset:
Aligns the STAT table in its home position.
G/Rotation:
Rotates the STAT table according to user specs. (CW, CCW, step 1 to 22)
H/Pulse:
Pulses the STAT table the number of steps specified from 1 to 999.
Rack Carry-in
I/Reset:
Resets the rack receiver unit.
J/Step:
Steps the rack as far as the rack buffer.
Rack Carry-in 2
L/Reset:
Rack holder moves to the right.
M/Step:
The rack push bar steps forward as far as the sampling belt and then back as the
Rev./Diag button is pressed again.