D-30
Chapter D - Software
Automated Chemistry Analyzer
AU2700 User's Guide
May, 2002
Set
Indicates if the STAT table is ready to load sample cups.
m
essaGe
m
eaninG
C
hange
in
baCkground
Color
SET
Ready to place sample cups
Blue
SET
Not ready to place sample cups
Grey
Checked/Checking/Unchecked
Checks the STAT table for errors.
m
essaGe
m
eaninG
c
hanGe
in
bacKGround
color
CHECKED
Check performed, no errors found
Blue
CHECKED
Check performed, errors found
Yellow
CHECKING
Check in progress, an error occurred Red
UNCHECKED Check not performed or check failed Red
STAT Calibration/QC
Indicates whether the calibration/QC is manual or automatic.
m
essaGe
b
acKGround
color
Manual
Grey
Automatic
Grey
Sample dispense
Indicates sample dispense status.
Circle Color
Red
Not dispensed. Patient samples were not analyzed because samples
were not requisitioned or reagent is missing.
Blue
Sample dispensed.
Yellow
Dispensing sample.
White
Cup detected.
Grey
No cup detected.
Ring Color
(Indicates sample type)
Blue
Reagent Blank
Yellow
Calibrator
Green
QC
Red
STAT
Orange
Repeat STAT
F5: Check Start:
Used to check the STAT table status. For procedures, refer to Chapter C, “Performing a
calibration or reagent blank on the STAT table,” or “Performing QC on the STAT table.”
F6: Cal Req. (Requisition calibration STAT analysis):
For procedures, refer to the “Performing Calibrations and a Reagent Blank” section of the Basic
Operations Chapter.
F7: QC Req. (Requisition QC STAT analysis):
For procedures, refer to the “Quality Control Analysis” section of the Basic Operations Chapter.