Automated Chemistry Analyzer
AU2700 User's Guide
Chapter D - Software
D-123
February 4, 2004
Use this software screen to:
Set individual test parameters:
1. Select [Parameters], [Calibration], [Calibration Specific].
2. Use the pull-down menu or <</>> buttons on screen to select a test name.
3. Select a sample type using the pull-down menu.
4. Press function key F4 (set). According to the messages displayed, set analysis
parameters.
5. Press function key F6 (Inner/Outer) to select the inner or outer cuvette wheel for MB
factors.
6. Press function key F2 (exit).
Parameters set in this process will be saved.
Set an ISE test:
Set the individual test parameters of each ISE test (Na, K, Cl). Perform the following using a
test number (97 for Soduim, 98 for Potassium and 99 for Chloride) in the [Calibration Specific]
screen.
1. Select [Parameters], [Calibration], [Calibration Specific].
2. Select the "ISE" tab located in the upper left corner of the screen.
3. Use the pull-down menu or </> buttons on screen to select a test name.
4. Select a sample type using the pull-down menu.
5. Press function key F4 (set). According to the messages displayed, set analysis
parameters.
Cal. type:
Set the calibration type for each analysis test to MCAL or ACAL. The default
setting is MCAL. Counts: 1 to 4 for ACAL only. Factor A: slope or multiplier, B: offset or
constant, cal. no., conc. factor range: for ACAL.
6. Press function key F2 (exit).
Parameters settings will be saved.