54 - Active Terminals and Connections
Voltage Generation
MC6-T has two voltage generation terminals as shown in
.
•
TC1
, generation range: -1 to +1 VDC.
•
OUT
, generation range: -3 to +24 VDC.
Note that you may simulate non-supported thermocouple signals using
TC1
port. Since you actually generate (milli)volts, you need a data table to
convert desired temperatures to millivolts.
Figure 40: Voltage generation terminals
.
Note:
It is advisable to enter 0 V output before connecting the
circuit.
Warning:
Short circuiting the voltage output may result in
damage to MC6-T and/or the connected instrument.
Resistance Simulation
Resistance simulation is available from R1 terminals only.
Use of 2-, 3- or 4-wire connection is up to the receiving instrument.
Connect the possible third and fourth wire according to the requirements of
the connected instrument, but
use only the two leftmost RTD1 terminals
in MC6-T
. See
Figure 41: Resistance simulation terminals. Range 0 …
.
MC6-T monitors the resistance measurement current. If the current is too
high, MC6-T cannot simulate the right resistance value and displays an
error message.