154 - Additional Information
Figure 147: Sensor Calculation Formula options
Note:
If you need to enter very small or big number into
Factor
field, open the menu available in
Soft Numeric Keypad
.
When applicable, the menu contains a possibility to insert an
exponent. This allows entering numbers in scientific notation,
e.g. 5.775E-07.
Callendar van Dusen Formula for PRTs
When
CvD RTD (R
0
, A, B, C)
, i.e. Callendar van Dusen formula is
selected, sensor coefficients can be added, see
configuration page for Callendar van Dusen constant and coefficients