User Defined Function Data - 159
Figure 152: Sensor test configuration and single point test page
The lower part of the first page allows you to define step sizes for tables
seen on the second and third page. There you can see the sensor's
temperature vs. resistance correlation (and vice versa) in table format.
User Defined Transfer Functions
Transfer Functions
are available in Calibrator's
Scaling Tool
and in
instrument definitions
found in Documenting Calibrator.
When creating/selecting a user defined transfer function, tap on the
Transfer Function
button and browse to the
User Transfer Function
page.
The configuration consists of two (or more) pages as shown in
Figure 154: 2nd configuration page
page is for general definitions and the second for entering known transfer
function points. If you enter more points than can be shown on one page,
a further page is automatically added.