Overtemperature Protection Test Mode
MC6-T is equipped with an additional safety feature called
Overtemperature Protection Test Mode
. It gives a possibility to make
sure that the power is disconnected from the Temperature Block in case it
heats up over the maximum setpoint temperature.
To enter the
Overtemperature Protection Test
mode select
Settings
from the Home view. Tap
Maintenance
button, select
Change module
data
option and provide the required PIN code:
5656
. After accepting the
changes the device is in test mode.
Figure 139: Overtemperature Protection Test Mode window
To test the feature, enable the temperature generation to maximum
setpoint specific for your model. If the protective mechanism works
properly, you will be prompted with a message shown in
Overtemperature Protection Test Mode prompt message
. The power will
be disconnected from the Temperature Block and the device will start
cooling. To return to the normal operating mode, switch the Mains switch
off and then back on, to clear the error condition.
CAUTION:
If the maximum setpoint temperature stabilizes
without an error message being shown, contact Beamex. Safe
use of MC6-T is no longer ensured.